How are Cleanrooms Classified?

Cleanrooms are classified according to the number and size of particles permitted per volume of air. We work with ISO, FED (Federal) and GMP cleanroom classifications.

The end-user determines the target classification for their cleanroom, towards which Airology can design and build a cleanroom with the appropriate technology.

ISO Standard Cleanrooms

The ISO 14644-1:2015 – ‘Classification of Air Cleanliness’ is the European standard for cleanroom classification, and is used throughout the UK and EU.

This sets out classes 1 – 9, with Class 1 being the cleanest and Class 9 having a similar air quality to any ordinary room.

The ISO Class stipulates the maximum allowable particles per cubic metre by size, as shown in Table 1 below. The cleaner the requirement, the more air changes per hour are required.

To Read This Table

This table shows that in an ISO 7 cleanroom, each cubic metre of air must contain:

  • Fewer than (<) 2 930 particles which are greater than or equal to (≥)5 micron
  • < 83 200 particles ≥1 micron
  • < 352 000 particles ≥0.5 micron

US Federal Standard 209 (FED STD 209E) offers equivalent classes which range from 1 – 100,000 in steps aligned with ISO Classes 3-8. The FED class number refers to the maximum allowable particles >/=0.5 microns (μm) per cubic foot.

To discuss cleanroom classification requirements further, speak to one of Airology’s experts today. Call: 01527 894 279

Table 1: ISO Classes of Air Cleanliness by Particle Concentration & FED STD Equivalents

For each ISO class, this table shows the maximum allowable no. of particles ≥ each size per m³.

ISO CLASS NUMBER

≥0.1 µm≥0.2 µm

≥0.3 µm

≥0.5 µm

≥1 µm

≥5 µm

FED STD 209E Equiv.

ISO CLASS 1

10ii

iv

iv

iv

iv

vviii

ISO CLASS 2

10024ii

10ii

iv

iv

vViii
ISO CLASS 3

1 000
23710235ii

iv

v1

ISO CLASS 4

10 0002 3701 02035283ii

v10

ISO CLASS 5

100 00023 70010 2003 520832

iv, v, vi

100

ISO CLASS 6

1 000 000237 000102 00035 2008 320293

1000

ISO CLASS 7

iii

iii

iii

352 00083 2002 930

10,000

ISO CLASS 8

iii

iii

iii

3 520 000832 00029 300100,000

ISO CLASS 9vii

iii

iii

iii

35 200 0008 320 000293 000viii

Notes to Table 1

i. All concentrations are cumulative, e.g. for ISO Class 5, the 10 200 particles shown at 0.3µm include particles 0.3µm or larger

ii. These limits require testing at large volume air samples. Sequential sampling may assist in air quality control.

iii. Concentration limits at this particle size are not applicable due to high particle concentration.

iv. Sampling and statistical limitations make classification requirements inappropriate for particles of this size.

v. Sample collection limitations for particles in low concentrations and sizes greater than 1µm make classification at this particle size inappropriate, due to potential particle losses in the sampling system.

vi. To specify this particle size in association with ISO Class 5, the macroparticle descriptor M (number of macro-particles per m3) may be adapted and used in conjunction with at least one other particle size.

vii. This class is only applicable while the cleanroom is in operation.

viii. There are no FED Class equivalents to ISO 1, 2 or 9.

GMP Cleanroom Classification

The Good Manufacturing Practices (GMP) facility grades for cleanrooms are in the range Grade A, Grade B, Grade C and Grade D. They stipulate ISO grades at rest and in operation, as well as necessary equipment, such as particle monitoring systems, and prohibited equipment, such as sinks and drains.

Table 2: GMP vs ISO At Rest & In Operation (EU GMP Annex 1. Rev. 2022)

For each grade, this table shows the maximum permitted number of particles ≥ each size per m³ at rest and in operation

GMP Grade

At Rest

≥0.5 µm

At Rest
≥5.0 µm
At Rest
ISO Class Equiv.
In Op.
≥0.5 µm

In Op.
≥5.0 µm

In Op. ISO Class Equiv.

A
3520

n/a

ISO 5

3 520

n/a

ISO 5

B

3520

n/a

ISO 5

352 000

2 930

ISO 7

C

352 000

2 930

ISO 7

3 520 000

29 300

ISO 8

D

3 520 000

29 300

ISO 8

n/a

n/a

Notes to Table 2

(i) Classification including particles ≥5.0µm is dependent on CCS (Contamination Control Strategy) or historical trends
(ii) End user’s risk assessment and routine data will establish operation limits for Grade D

More about cleanroom classification:

Regulating the concentration of airborne particles prevents contamination by minimising particles and microbes entering, remaining and multiplying within the space. Cleanrooms may also control other environmental factors like temperature, humidity, and light levels as needed.

Relationship between Contamination and Cleanroom Classification

Cleanrooms must be designed, constructed, and operated according to the cleanliness classification defined in ISO 14644-1:2015.
The classification required depends on the specific tasks performed within that cleanroom. Products that are highly susceptible to contamination need cleaner spaces. Once a classification is determined, the cleanroom must be maintained to meet the appropriate specifications for environmental factors such as:

  1. Cleanliness
  2. Temperature
  3. Humidity
  4. Pressure
  5. Number of air changes per hour
  6. Flow rate

When a cleanroom is constructed in a pristine environment using non-particulate materials, airflow is the primary influence on its cleanliness. Depending on the ISO class, a cleanroom will require between 5-580 air changes per hour to keep particulate levels consistently low.

Impact of Particle Size on Classification

The ISO class required depends greatly on the size of allowable particulate matter. This is the diameter of the particle measured in microns or micrometres (µm). Particulate matter is sometimes abbreviated to PM, with the diameter indicated numerically, e.g. PM5 refers to particulate matter greater than or equal to 5µm across.

For example: if a customer requires a cleanroom with fewer than 1000 particles per m3, they may opt for an ISO 6 cleanroom, as this requires less than 293 PM5/m3. However, ISO 6 allows up to 8,320 particles of 1 micron across, and 1,000,000 particles of 0.1 micron across (PM0.1).

Mistakenly classifying based on the wrong particle size would lead to inadequate cleanliness, so it is important to know what sizes of particles would cause contamination in your application.

Similarly, mistakenly identifying particle sizes when verifying cleanroom classification can cause errors. Mistaking 0.5-micron particles for 5-micron particles could mean the difference between ISO 5 and ISO 8, so the decimal place on an air quality meter is crucial to correctly classifying spaces.

Positive vs Negative Pressure in Cleanrooms

Most cleanrooms use positive pressure to keep contaminants from entering the room.

  1. Air is introduced through a fan-powered filter to remove particles, creating a slightly higher air pressure inside the cleanroom than outside.
  2. The air and any contaminants generated within the room are pushed down to the floor level, where vents allow them to flow out to the exterior.
  3. The pressure keeps most contaminants from entering through vents or when the door is open.

Negative-pressure cleanrooms are used to isolate substances, fumes or particles created during internal processes such as medical research or pharmaceutical development. This protects the space outside the enclosure.

  1. A fan pulls air and contaminants out of the room through a filter.
  2. This creates negative pressure inside the room, drawing new air in through venting.
  3. The force of air entering the room prevents contaminants from escaping.

Negative-pressure cleanrooms usually form part of a layout with multiple cleanrooms, with an anteroom immediately outside that is maintained at a higher pressure. This allows researchers to enter and exit without disrupting the environment.

Independent Cleanroom Classification

We require independent inspections and testing before use to verify that the cleanrooms we build are operating at the required classification. Read more about this on our Cleanroom Validation information page.

Our commitment to quality is evident in every project we undertake – delivering innovative class leading solutions for all our projects..

We work closely with clients from every sector of industry including pharmaceutical, medical, veterinary, catering, research, engineering, and technology fields.

Airology can provide a complete bespoke turnkey solution for your project, from concept to completion, focusing on the design, manufacture and construction of innovative cleanroom and controlled environment solutions.

7 key decisions when creating a cleanroom environment

1. Decide how it’s to be used

2. Decide how it should be classified

3. Decide on the right environment

4. Decide what size it should be

5. Decide how impurities will be extracted or filtered

6. Decide on the right lighting condition

7. Decide how secure your facility needs to be

What our customers say…

"I have worked with Airology for several years and always found them to be reliable and responsive in regard to the servicing and maintenance of our HVAC equipment."

Colin Welsh

Orbit Group

"Throughout the project Airology provided us with guidance and their expertise to ensure that the installation met our specific requirements and was completed on time. We are delighted with our state of the art Clean Room and the professional approach and support given to us by the Airology team was invaluable."

Andrew Stringer

Managing Director, Entaco Ltd

"During the initial design and build we worked closely with Airology to create a facility that will provide one of the most comprehensive GMP stability storage and testing services in Europe. Working with Airology as the principal contractor enabled us to deliver a significant building project, on-scope, on-time and on-budget."

Mark Hammond

Managing Director, Intertek Melbourn Scientific

"Airology completed our installation with a high degree of professionalism and knowledge. P3 would have no hesitation in recommending the company for similar projects."

Simon Talbot

Managing Director, P3 Medical Ltd

"Great company to work with. Our cleanroom was designed and constructed by Airology and they still service and validate for us. Would definitely recommend"

Helena Flowers

Andel Plastics Ltd

"We engaged Airology for modifications in our Clean Room at our pharmaceutical company. The communication with their technical team was exceptionally effective. The service provided was of the highest quality, making them an excellent choice for conducting business."

Suzani Menegon

SyriMed Ltd

"Airology have recently designed and built our cleanroom. They were a pleasure to work with and provided a very professional service. I would have not hesitation in recommending the team."

David Shand

MatOrtho Ltd

"Airology have always been professional and helpful when we have needed advice at Insight Medical Products with our Cleanroom. Thoroughly recommend."

Stacey Spruels

Insight Medical Products Ltd

"Professional company, very friendly and experienced team. They designed for our pharmaceutical site the best flexible and efficient HVAC considering practicality, potential to expand and costing. I would recommend working with this company. They are very transparent and keen to solve any difficulties during the project."

Zakaria Saad

Medellux Ltd